Skip to content
  • There are no suggestions because the search field is empty.

Connecting to a RIEGL scanner using an ethernet cable

Sometimes it is necessary to connect to a scanner over ethernet; for example, to download scan data, log files, or to perform a firmware update.

This guide focusses on terrestrial (TLS) scanners, but the steps are similar for mobile and airborne scanners. You will need to be able to configure the IP settings of the PC you want to connect from, along with some other networking settings, and this might require admin access to Windows.

The simplest method to connect to the scanner is to use static IP addresses on both the scanner and the PC. Most RIEGL TLS scanners will be set to a static IP address by default, but you can check this by reviewing the network settings on the scanner using the scanner's HMI (human-machine interface).

Scanner settings

On a VZ-i series scanner, tap the network status icon at the top of the screen:

Check the "Mode" is set to "Static IP". If not, tap the purple text to change it:

Make a note of the IP Address - normally it will be 192.168.0.129.

If you have made any changes, tap "Apply" and wait for the changes to be saved.

Connect the green ethernet cable that was supplied with the scanner to the ethernet port on the base of the scanner. Take care not to bend the pins! Line up the notch in the socket with the peg in the plug on the cable. Sometimes the thread can bind up as you screw it in - if this happens, gently push the cable into the socket every quarter-turn to unbind the thread, and you will be able to turn it again.

PC settings

Plug the RJ45 end of the ethernet cable into the ethernet port on the PC. Once the cable is connected, the green status lights on the PC port should illuminate (though they are not always present on laptops with space-saving ethernet ports). This indicates am electrical connection, but doesn't necessarily indicate that the port is configured correctly yet top allow communication.

Open the Windows Ethernet settings page (you can simply search for "ethernet" in the Windows search box).

Check the "IP assignment" setting:

 You want it to be set to "Manual" - if it isn't, click "Edit", and configure it as shown:

  • First drop-down: Manual

  • IPv4: On

  • IP address: 192.168.0.100

  • Subnet mask: 255.255.255.0

  • The remaining fields can be left as default

Click "Save". Check the IP assignment settings match these:

You should now have an ethernet connection established between the PC and the scanner.

Test the connection

To check the connection, you can "ping" the scanner from a Windows command prompt:

Open the Command Prompt (search for "cmd" in the Windows search box) and type following command:

ping 192.168.0.129

Press Enter and look for the following confirmation: